Selecting The Input Impedance And Input Coupling - YOKOGAWA TA220 User Manual

Digital jitter meter
Table of Contents

Advertisement

4.2
Selecting the Input Impedance and Input
Coupling
Procedure
KEY
LOCK
SEC RANGE
Explanation
IM 704610-01E
Display 2
Display 1
ns ms
s
%
V
%/
S
Selecting the Input Impedance
Press IMPEDANCE, and select 50 Ω or 1 MΩ. The indicator of the selected
item illuminates.
Selecting the Input Coupling
1.
Press SHIFT+IMPEDANCE (COUPLE). Couple appears in Display 2, and dC
or AC appears in Display 1.
2.
Turn the rotary knob to select dC or AC.
Input Impedance
You can select an input impedance for the measurement input terminal of 50 Ω or 1 MΩ.
Input Coupling
You can select an input coupling for the measurement input terminal of DC or AC.
Note
The frequency characteristics differ depending on the input impedance and input coupling
setting. For information on the frequency characteristics, see section 13.1.
<<For a functional description, see section 2.4>>
L.MARK
R.MARK
DISP OFF
MEASURE
E2T
JUDGE
PW
GO
BD x1
D to C
NO-GO
GATE TIME
DELAY
BLOCK
FUNCTION
MEDIA
ARMING
INHIBIT
SHIFT
SHIFT
REMOTE
STORE
UTILITY
10%
LOCAL
RECALL
ENTER
20%
LEVEL
JDG LEVEL
COUPLE
EQ BOOST
S
POLARITY
TRIG
IMPEDANCE EQUALIZER
50 Ω
AUTO
MAN
1M Ω
CONV
SEC RANGE
SCALE
The phrase, "rotary knob & <>"
is used in the following
explanation to instruct the user
to enter numerical setting values
using the knob and arrow keys.
For details on this procedure,
see section 3.7.