Chapter 5: Troubleshooting
To test the cables

Adjust sampling positions using Eye Finder

1 Select the Sample Positions button. The Eye Finder window will appear.
2 In the "Buses/Signals" section of the Eye Finder window, ensure that the
check box next to "My Bus 1" is checked.
3 Drag the blue bar for "My Bus 1" to approximately -3 ns.
4 Select the plus sign to expand bus "My Bus 1".
5 Select the Run button in the Eye Finder window.
6 Ensure that an eye appears for each bit. Depending on your test setup, the
eye position may vary. Ensure that all blue bars in the individual channel
rows (excluding the top row) are to the left of the orange transition