Automatic Blank (Non-linear) Correction Feature
response during a multipoint calibration. A separate blank does not have to be run. Graphically, automatic blank correction
E = Eo + S * log (C + b)
E1 = Eo + S * log (C1 + b)
E2 = Eo + S x log (C2 + b)
E3 = Eo + S x log (C3 + b)
the multipoint calibration is treated independently.
Thermo Scientific Orion Star A214 Benchtop and Star A324 Portable pH/ISE Meter Reference Guide