Save Full Backup (Presets, Configurations & Application Settings) - GE Voluson 730Pro Service Manual

Hide thumbs Also See for Voluson 730Pro:
Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
4-5-3
Save Full Backup (Presets, Configurations & Application Settings)
A full backup always contains the following data
Patient demographic and exam data (database containing the patient data and measurements)
SonoView image data (NOT available when saving to the internal hard disk, DVD/CD or MOD)
User Settings (databases and files containing gray curves and the user settings.)
Image transfer settings (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
Measure Setup Settings (user specific measure settings)
V730 settings (general settings such as language, time/date format and the enabled options)
Windows Network Settings (network settings including the computer name)
Service Platform (state of the service platform)
VP (additional system data)
It is recommended to create a full backup of the settings once a week.
CAUTION
!! CAUTION:
NOTE:
Always "Full Backup" any presets, configurations and application settings to HDD and/or DVD or MO-
disk before upgrading the software and/or application settings. This ensures that if the presets need to
be reloaded, will be the same ones the customer was using prior to service.
1.) Press the UTILITIES key on the control panel. The menu area changes to the Utilities menu.
2.) Select the SYSTEM item from the menu area to activate the setup desktop screen.
3.) Select the
4.) Choose the Destination (e.g., Network Drive).
5.) Enter the description of the full backup.
6.) If desired and possible (Network Drive and Other drive only), activate "Include Images.
NOTE:
The "Include Images" option may result in a large amount of data: up to 70 Gigabytes!
7.) Select the NEXT button to start the backup process.
After copying the data, the Voluson® 730Pro / ProV reboots and the application starts again.
4-34
Section 4-5 - Backup and Restore Database, Preset Configurations and Images
DRAFT (A
page and click the SAVE button of the "Full Backup" group.
BACKUP
Figure 4-25 Full Backup Save
28, 2008)
PRIL
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
choose destination
enter description
click [Next]