GE Voluson 730Pro Service Manual page 117

Hide thumbs Also See for Voluson 730Pro:
Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
Table 4-6
Step
11
DISPLAY MODE
12
SCALE (CFM, TD)
13
LINE FILTER
14
GENTLY COLOR
15
ARTEFACT (on/off)
16
BASELINE
17
ENSEMBLE
18
LINE D.
19
C-MAP (PD, TD)
20
SM. RAISE
21
SM. FALL
22
FLOW RES.
23
BALANCE
24
THRESHOLD
Exit
25
For further details refer to the Voluson® 730Pro / ProV Basic User Manual:
Chapter 8, CFM Mode (Color Flow Mode)
Chapter 9, PD Mode (Power Doppler Mode)
Chapter 10, TD Mode (Tissue Doppler Mode)
DRAFT (A
Color Doppler Mode (CFM, PD, TD) Functions
Task
To select the CFM- Display Mode (V; V-T; V-Pow; Pow-T; or T).
The maximum velocities are displayed above and under the color scale in kHz, cm/s or m/s.
With "Line Filter" the signals of the neighboring pulses are less weighted for the display of the
actual pulse which considerably improves the detail lateral resolution and signal-to-noise ratio.
Gently means the transition between color and gray scale information.
The embedding of the color into 2D Mode is performed smoothly with less colored splashes.
Switch on/off the artifact suppression.
The baseline shift can be used to prevent aliasing in one flow direction similar to the Doppler
baseline shift. There are 8 steps in each direction. (impossible in PD Mode)
Controls the number of pulses to constitute one Color- or Power-Doppler line in the display.
Determines the line density within the Color-Box.
The lower the line density, the larger the line distance and the size of the color pixels.
Provides selectability of the color coding for an optimization of the display of blood flow (similar
to the post-processing curves with grayscale 2D scans).
After a selection has been made, the color bar displays the resultant map.
To select different filter periods for raising velocity.
Filtering of the rise velocity leads to noise suppression.
To select different filter periods for falling velocity.
This filter leads for "prolongation" of the display flow.
The "Flow Resolution" function controls the axial resolution of color in the display.
It adjusts the axial sample depth of color pixels.
The Balance controls the amount of Color display over bright echoes and helps to confine
color within the vessels wall.
After FREEZE you can adjust the Color Threshold. It eliminates small color noise or motion
artifact signals in the color image. (small number cuts off fewer signals than a higher setting)
Press the EXIT key on the control panel to exit the CFM, PD or TD Sub menu.
Chapter 4 - Functional Checks
28, 2008)
PRIL
Expected Results
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
V (BT03)
M
ANUAL
4-15