Probe Maintenance

10-5-7-1
Probe Related Checks
Table 10-10 Probe Related Checks
Step
1
2
10-5-7-2
Basic Probe Care
The Basic User Manual provides a complete description of probe care, maintenance, cleaning and
disinfection. Ensure that you are completely familiar with the proper care of GE probes.
Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of
Voluson® 730Pro / ProV for more details. Failure to follow these precautions can result in serious injury
and equipment damage. Failure to properly handle or maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and system sockets before plugging in a probe.
10-5-7-3
Basic Probe Cleaning
Refer to the Basic User Manual of Voluson® 730Pro / ProV for details on cleaning.
Failure to follow the prescribed cleaning or disinfection procedures will void the probe's
CAUTION
warranty. DO NOT soak or wipe the lens with any product not listed in the Voluson® 730Pro /
!! CAUTION:
ProV Basic User Manual and/or care card. Doing so could result in irreparable damage to the
probe and/or Voluson® 730Pro / ProV system.
Follow the Care Card instructions supplied with each probe (inside the transducer boxes)
CAUTION
for disinfectants and gels that are compatible with the surface material of the probes.
!! CAUTION:
To help protect yourself from blood borne diseases, when cleaning and handling probes,
CAUTION
wear approved, non-allergenic disposable gloves.
!! CAUTION:
Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.
NOTICE
!! NOTICE:
CAUTION
! CAUTION:
Please be aware of the
sensitive probe head.
TAKE EXTREME CARE!
10-10
DRAFT (A
Item
Thoroughly check the system probe connectors and remove dust from inside the connector
Probes
sockets if necessary. Visually check for bent, damaged or missing pins.
Probe Holder
Clean probe holders (they may need to be soaked to remove excess gel).
NEVER place or store a
probe on its scan head!
Section 10-5 - System Maintenance
28, 2008)
PRIL
Description
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
When disinfecting a probe,
ensure that there is sufficient
space between the probe and
the container bottom!
V (BT03)
M
ANUAL

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson 730prov

Table of Contents