GE Voluson 730Pro Service Manual page 236

Hide thumbs Also See for Voluson 730Pro:
Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
M_Gain Out Of Range
M_Reject Out Of Range
MC Ensemble Out Of Range
MC_Balance Out Of Range
MC_BaseLinePos Out Of Range
MC_Dynamic Out Of Range
MC_FallSmooth Out Of Range
MC_Gain Out Of Range
MC_RiseSmooth Out Of Range
M-DynContrast Out Of Range+
Memory access violation
memory allocation error
missing BCMCPW_TxFocusData
missing BM_TxFocusData
missing CW_TxFocusData
missing ProbeAcousticUnit, wrong ProbeAcousticUnitID.
missing ProbeGeneral data - wrong ProbeID
missing ProbeScanFunc - wrong ProbeScanFuncID
MotCtrl RS232 communication timeout
MotCtrl: No Referenzposition signal!
MotionColor-DSC 1st:Memory allocation failed
MotionColor-DSC 2nd:Memory allocation failed
Motion-DSC:Memory allocation failed
MoveVolumeAxis only supported by RealTime-probes !
MoveVolumeAxis only supported by RealTime-probes (RRE)
No CD Writer found
No disc in drive
Not enough space. (Not enough space on destination to hold
the backup data.)
7-26
DRAFT (A
PRIL
Error Messages
Section 7-8 - Error Messages
V
OLUSON
28, 2008)
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
press ok and save this user-setting once again
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
Confirm the message by pressing the OK button,
disconnect and reconnect the active probe and use it
again. If the message appears again, usually the probe
itself has an defect, so you should contact the service
restart the system
restart the system
restart the system
look into the systemSetup to see the current MotCtrl
version and contact technical support
look into the systemSetup to see the current MotCtrl
version and contact technical support
check the connection and the Power cable - plug the cable
off and on and try again. (restart the system)
insert disk, if fails again reboot and try again (with another
disk)
select another destination to save Full Backup
® 730P
/ P
V (BT03)
RO
RO
S
M
ERVICE
ANUAL