Section 7-5
How to use the Auto Tester program
The following message box apppears.
5.) Click OK.
6.) Press the PAUSE key on the keyboard.
The Auto Tester window appears, see:
Please check status of the "Accumulate mouse moves" function. It has to be turned OFF.
NOTICE
i.e., no check mark visible!
7.) Activate the "Auto Tester" program by clicking the "Record" icon on the displayed screen.
8.) Start scanning.
You can scan normally and everything will be recorded to the program (up to several hours.)
NOTE:
It is important that you are recording the processes where the errors normally occur.
9.) Stop the program by pressing the PAUSE key on the right upper corner of the keyboard.
7-8
DRAFT (A
Figure 7-6 Message Box
Figure 7-7
Figure 7-7 Start Auto Tester
Section 7-5 - How to use the Auto Tester program
28, 2008)
PRIL
(cont'd)
below.
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
Pause
key on the
right upper corner
of the keyboard
"Record"
icon
"Accumulate mouse
moves" function has
to be turned OFF
(no check mark)
V (BT03)
M
ANUAL