Section 8-8
Replacement or Activation of Options
Following Options are available:
•
Real Time 4D
•
DICOM
•
CRI (Compound Resolution Imaging)*
•
VOCAL*
•
RT_4D_Biopsy*
NOTE:
Additional option fields are not yet implemented in the Voluson® 730Pro / ProV.
Not all Options are available on the Voluson® 730Pro V (marked with * in this manual).
NOTICE
!! NOTICE:
8-8-1
Operation for activating Options
1.) Press the UTILITIES key on the control panel once to display the Utilities menu.
2.) Select the SYSTEM item from the menu area to invoke the setup desktop on the screen.
3.) Select the
8-8-1-1
Operation for installing an Option:
1.) Position and select the option input field you want with the write cursor.
2.) Clear/Edit the current number, if existing.
3.) Enter the encrypted serial code or option code with the keyboard.
4.) Click the SAVE&EXIT button.
(The code will be checked and if ok the serial number/option is accepted.)
NOTE:
After installation of the DICOM option, restart (turn off an on) the Voluson® 730Pro / ProV system.
DRAFT (A
page where you can see which options are installed in the system.
OPTION
Option page of
System type:
Voluson 730Pro
Figure 8-21 Option page in the System Setup
Chapter 8 - Replacement Procedures
28, 2008)
PRIL
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
Option page of
System type:
Voluson 730ProV
V (BT03)
M
ANUAL
8-15