Section 4-3

General Procedure

SYSTEM REQUIRES ALL COVERS
CAUTION
Operate this unit only when all board covers and frame panels are securely in place.
!! CAUTION:
The covers are required for safe operation, good system performance and cooling purposes.
Lockout/Tagout Requirements (For USA only)
NOTICE
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the Power Cable
!! NOTICE:
on the system.
TAG
&
LOCKOUT
Signed
Date
4-3-1
Power On / Boot Up
NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-1-1
Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Screw on the pull-out protection of the mains power cable with the 2 screws.
3.) Connect the Power Cable to an appropriate mains power outlet.
4.) Switch ON the Circuit Breaker at the rear of the system.
Circuit Breaker
Mains Power Cable
Figure 4-1 Circuit Breaker, Protection and Power Cable on back of Voluson® 730Pro / ProV
When AC power is applied to the scanner, the ON/OFF switch on the Control panel illuminates amber,
NOTICE
indicating the System (including the Back-end Processor) is in standby mode.
!! NOTICE: