GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
Section 8-12
Replacement of the Trackball top fixation ring
8-12-1
Manpower
One person, 5 min.
8-12-2
Trackball top fixation ring - Replacement Procedure
1.) Remove the fixation ring by turning it counterclockwise.
rotate clockwise to
mount the fixing ring
2.) Mount the fixation ring by turning it clockwise.
Section 8-13
Replacement of Digipots and TGC Slider controls
8-13-1
Manpower
One person, 5 min.
8-13-2
Tools
small-sized slotted screwdriver or tweezers
8-13-3
Cap and/or Spring - Replacement Procedure
1.) Remove the cap (for Rotation digipots or for Slider-potentiometer TGC).
2.) Mount the cap (for Rotation digipots or for Slider-potentiometer TGC).
DRAFT (A
PRIL
Figure 8-29 Trackball with top fixation ring
Chapter 8 - Replacement Procedures
V
OLUSON
28, 2008)
rotate counterclockwise
to remove the fixing ring
® 730P
/ P
V (BT03)
RO
RO
S
M
ERVICE
ANUAL
8-23