From Software Version 3.1.x onwards, it is also possible to save/load/delete "Full Backup" data via the
NOTICE
System Setup
!! NOTICE:
1.) Press the UTILITIES key on the control panel. The menu area changes to the Utilities menu.
2.) Select the SYSTEM item from the menu area to activate the setup desktop screen.
3.) Select the
"User Settings Only"
group
The "Backup" page is subdivided in two main groups:
A.) User Settings Only
•
Section 4-5-1 "Save User Settings Only (Application Settings)" on page 4-31
•
Section 4-5-2 "Load User Settings Only (Application Settings)" on page 4-32
B.) Full Backup
•
Section 4-5-3 "Save Full Backup (Presets, Configurations & Application Settings)" on page 4-34
•
Section 4-5-4 "Load Full Backup (Presets, Configurations & Application Settings)" on page 4-36
•
Section 4-5-5 "Delete Full Backup (Presets, Configurations & Application Settings)" on page 4-38
The User Settings and/or Full Backup can be saved to the following destinations:
•
D: partition of internal hard disk
•
DVD/CD+(R)W
•
MOD (if present)
•
Mapped Network Drive Z:
see:
•
Any other drive connected to the system (e.g.; an external USB-hard disk)
Note: This function is only available in the Full Backup utility.
For further details review:
When connecting external USB devices, be sure to execute Safety Directions found in Chapter 2 of the
NOTICE
Voluson® 730Pro / ProV (BT03) Basic User Manual.
!! NOTICE:
4-30
Section 4-5 - Backup and Restore Database, Preset Configurations and Images
DRAFT (A
page; see:
BACKUP
page.
BACKUP
Figure 4-20 System Setup - Backup page
Section 3-11-1 "Map Network Drive" on page 3-49
Section 3-5-11 "External USB-Devices" on page