Power Off / Shutdown - GE Voluson 730Pro Service Manual

Hide thumbs Also See for Voluson 730Pro:
Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
4-3-1-1
Scanner Power On (cont'd)
5.) Press the ON/OFF Standby switch left below the Control Panel.
ON/OFF Standby Switch
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
When the ON/OFF Stand-By switch left below the Control Panel is pressed, the System (including the
Back-end Processor) starts and the software code is distributed to initiate the scanner.
As soon as the software has been loaded, the system enters 2D-Mode with the probe and application
that were used before the system was shut down. Depending on the BIOS-Version no status messages
are displayed during this process. Boot up time is about 2 minutes.
4-3-2

Power Off / Shutdown

After turning off a system, wait at least 10 seconds before turning it on again.
NOTICE
The system may not be able to boot if power is recycled too quickly.
!! NOTICE:
4-3-2-1
Scanner Shutdown
1.) Press the ON/OFF Standby switch left below the Control Panel.
2.) Switch OFF the Circuit Breaker at the rear of the system.
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
Disconnection of the Main Power Cable is necessary!
WARNING
WARNING
For Example: When repairing the system.
!! WARNING:
3.) After complete power down, unscrew the 2 screws and remove the pull-out protection to disconnect
the main power cable from the system or unplug it from the AC wall outlet socket.
Refer to
4.) Press once on the brakes to block the front wheels (brakes on front wheels under the foot rest).
5.) Disconnect probes. (Turn the probe locking handle counterclockwise and then pull the connector
straight out of the probe port.)
DO NOT disconnect a probe while running (Live Scan "Write" mode)!
CAUTION
A software error may occur. In this case switch the unit OFF (perform a reset).
!! CAUTION:
DRAFT (A
Figure 4-2 ON/OFF Standby Switch
Figure 4-1 on page
4-2.
Chapter 4 - Functional Checks
28, 2008)
PRIL
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
V (BT03)
M
ANUAL
4-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson 730prov

Table of Contents