Section 3-5
Connection of Auxiliary Devices
3-5-1
Preparations
NOTE:
Normally the auxiliary devices and peripherals come already installed with the system.
1.) Carefully remove plastic caps using a knife and loosen screws and washers (see:
2.) When the cover is loose on top, pull the rear cover out and move upwards.
plastic caps; screws and washers
3.) Disconnect the GND ground-cable from the back of the rear cover plate.
4.) Connect Peripherals according to correct connection scheme described in the following sub-
sections (refer to
5.) Connect the GND ground-cable at the rear cover plate.
6.) Mount rear cover plate, reattach screw and washers and plug-in the caps.
3-8
DRAFT (A
PRIL
Figure 3-5 Remove rear cover plate
Figure 3-6 GND ground-cable
Table 3-6 on page
3-9).
Section 3-5 - Connection of Auxiliary Devices
V
OLUSON
28, 2008)
® 730P
/ P
V (BT03)
RO
RO
S
M
ERVICE
ANUAL
Figure
3-5).
GND ground cable