Save Full Backup (Presets, Configurations & Application Settings); Full Backup Save - GE Voluson 730Pro Service Manual

Hide thumbs Also See for Voluson 730Pro:
Table of Contents

Advertisement

GE M
S
- K
EDICAL
YSTEMS
D
KTI105947, R
IRECTION
4-5-3
Save Full Backup (Presets, Configurations & Application Settings)
A full backup always contains the following data
Patient demographic and exam data (database containing the patient data and measurements)
SonoView image data (NOT available when saving to the internal hard disk, DVD/CD or MOD)
User Settings (databases and files containing gray curves and the user settings.)
Image transfer settings (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
Measure Setup Settings (user specific measure settings)
V730 settings (general settings such as language, time/date format and the enabled options)
Windows Network Settings (network settings including the computer name)
Service Platform (state of the service platform)
VP (additional system data)
It is recommended to create a full backup of the settings once a week.
CAUTION
NOTE:
Always "Full Backup" any presets, configurations and application settings to HDD and/or DVD
or MO-disk before upgrading the software and/or application settings. This ensures that if the
presets need to be reloaded, will be the same ones the customer was using prior to service.
1.) Press the UTILITIES key on the control panel. The menu area changes to the Utilities menu.
2.) Select the SYSTEM item from the menu area to activate the setup desktop screen.
3.) Select the
4.) Choose the Destination (e.g., Network Drive).
5.) Enter the description of the full backup.
6.) If desired and possible (Network Drive and Other drive only), activate "Include Images.
NOTE:
The "Include Images" option may result in a large amount of data: up to 70 Gigabytes!
7.) Select the NEXT button to start the backup process.
After copying the data, the Voluson® 730Pro / 730ProV reboots and the application starts again.
RETZTECHNIK
2
EVISION
page and click the SAVE button of the "Full Backup" group.
BACKUP
Figure 4-27 Full Backup Save
Chapter 4 - Functional Checks
DRAFT
V
OLUSON
® 730P
/ 730P
V (BT´04)
RO
RO
S
M
ERVICE
choose destination
enter description
click Next
ANUAL
4-35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson 730prov

Table of Contents