GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
5-11-2-3

Fuses on Rear Panel (CPN80-81)

F1 + F2: The main Input voltage is fused by two 16 Ampere fuses (rated current 16A) labeled F1 and F2.
F3: The AC Output voltage (115/230V) is fused by either:
• a 1.6 Ampere / 250V fuse for auxiliary output voltage 230 VAC
• a 3.2 Ampere / 250V fuse for auxiliary output voltage 115 VAC
F4: 16 Ampere fuse for the input voltage for the switching power supply generating the DC-Supply
voltage for the Secondary power supply inside GEF-Module
5-11-2-4
Fuses inside CPN80-81
F1 on CCF board: fuses the surge current limiter circuit.
NOTE:
If this fuse is blown, the NTC (limiting the surge current will remain hot during system operation and if
the system is switched off/on within a few seconds the surge current could be to high.
Because of this reason Fuse F1 on CPN Rear Panel or the Hospital circuit fuse could be blown.
5-11-3
Disk Drive Module (GEM)
The check points for the drives voltages can be found at the AUX Supply connector (= GEM Power
connector) on Backpanel of GEF-Box; see:
MO drive is supplied by +5V only
DVD/CD drive is supplied by both +5V and +12 V
Fan on the GEM module needs only 12 V power.
5-11-3-1
Fuses on CPE-Board for Disk Drive Module (GEM)
Description of fuses: 5A / 250V / slow-blow fuse / 5 x 20 mm
F1
+12V
5-44
DRAFT (A
28, 2008)
PRIL
Figure 5-27 Fuses of CPN80-81
Section 7-3 "Check Points Voltages" on page
Figure 5-28 fuses for disk drives
Section 5-11 - Power Distribution
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
7-4.
V (BT03)
M
ANUAL
F2
+5V

Advertisement