Backup System Presets And Configurations - GE LOGIQ 9 Technical Manual

Hide thumbs Also See for LOGIQ 9:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2294854-100, R
IRECTION
4-3-11-2

Backup System Presets and Configurations

NOTE:
Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
1.) Insert a formatted MOD (Magneto-Optical Disk) or CD-R into the drive.
2.) On the Touch Panel, press UTILITY.
3.) On the Utility Touch Panel, press SYSTEM.
4.) On the monitor display, select BACKUP/RESTORE.
NOTE:
If you are not logged in as GE Service or with administrator privileges, the Operator Login
window is displayed. Log on with administrator privileges.
5.) In the Backup list, select Patient Archive, Report Archive, System Configuration and User Defined
Configuration.
6.) In the Media field, select MO or CD-RW.
7.) Select BACKUP.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.
Check here to
backup presets
and configurations
4-16
PROPRIETARY TO GE
3
EVISION
Figure 4-13 Backup/Restore Menu