Verify the following:
6.
Active audio alarm.
a.
Flashing %SpO
b.
BPM indication in the range of 57 to 63.
c.
Pulse waveform of approximately 1-inch P-T-P amplitude.
d.
Press the SRC-MAX %SpO
7.
system displays three dashes [ - - - ] until the %SpO
Service Manual
Figure 4-35.  BPM of 60 with High Modulation
indication in the range of 73 to 77.
2
selection button. The SRC-MAX %SpO
2
Operational and Functional Tests
90 LED lights. The monitoring
2
stabilizes to a value in the range of 88 to 92.