Modification and Testing
Verify the following:
11.
The SpO
The High priority alarm sounds.
The message "Sensor Off" appears.
Screen Saver
Screen Saver saves power by darkening the screen.
The Screen Saver functions during periods when no alarm condition exists and the buttons are
not pressed.
To set the Screen Saver
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
1.
Access the Service Menu.
2.
In the Service Menu, select the Power Saving Settings menu.
3.
Select Screen Saver Time. The factory default is 3 minutes.
4.
Set the desired number of minutes.
5.
Screen Saver Time selections include Never and from 1 minute to 10 minutes.
In the Power Saving Settings menu, select the Screen Saver Brightness menu.
6.
4-24
and pulse rate values are highlighted by flashing red boxes.
2
Figure 4-25.  Screen Saver Time Setting