Service Menu Options; Overview; Safety Reminders; Service Menu Overview - Covidien Nellcor PM10N Service Manual

Portable spo2 patient monitoring system
Hide thumbs Also See for Nellcor PM10N:
Table of Contents

Advertisement

2 Service Menu Options

Overview

2.1
This chapter describes Service Menu settings for the Nellcor™ portable SpO
system. A password is required to access the Service Menu. Contact Technical Services
(reference
Technical

Safety Reminders

2.2
WARNING:
For best product performance and measurement accuracy, use only accessories supplied or
recommended by Covidien. Use accessories according to their respective Instructions for Use.
WARNING:
The sensor disconnect error message and associated alarm indicate the pulse oximetry sensor is
either disconnected or has faulty wiring. Check the connection and, if necessary, replace the
sensor, the pulse oximetry cable, or both.
WARNING:
Do not use damaged pulse oximetry sensors. Do not use with exposed optical components. Do
not immerse completely in water, solvents, or cleaning solutions, since pulse oximetry sensors
and connectors are not waterproof. Do not sterilize by irradiation, steam or ethylene oxide. Refer
to the cleaning instructions in the Instructions for Use for reusable sensors.
Caution:
Do not attach any cable intended for computer use to the sensor port connector.

Service Menu Overview

2.3
WARNING:
Only qualified service personnel should open the monitoring system housing, remove and
replace parts, or make adjustments.
Services, page 1-8) for the default password.
patient monitoring
2
2-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents