Modification and Testing
Press the Down button to decrease the brightness.
•
Press the Up button to increase the brightness.
•
Press OK to save the desired brightness.
3.
Save Spot Reading
The Save Spot Reading function saves a point in time of the patient's data.
To save a spot reading
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system off.
1.
Connect the DEC-4 pulse oximetry cable to the monitoring system sensor port.
2.
Connect the OxiMax™ DS-100A sensor to the DEC-4 cable, then to a finger.
3.
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
4.
Press Menu.
5.
Highlight Save Spot Reading and press OK.
6.
The message "Spot Reading Saved" appears.
4-20
Figure 4-20. Brightness Setting
Service Manual