Modification and Testing
Exit the Service Menu.
9.
As soon as the SpO
10.
immediately silence the alarm tone.
With the alarm silenced, verify the following:
11.
Alarm remains silent for 30 seconds before the alarm tone is audible.
Alarm Silenced indicator lights, and the Alarm Silenced message appears in the display.
SpO
2
The pulse tone is audible.
Permission to Mute Alarm
Use the same values for Low SpO
1.
Confirm the following results:
2.
The pulse tone is audible.
The SpO
The audible alarm sounds.
The following messages alternately appear in the display: Pulse Rate Low and SpO
In the Service Menu, select Permission to Mute Alarm.
3.
4-14
Figure 4-14.  Alarm Silence Duration Setting of 30 Seconds
Low and Pulse Rate Low alarms sound, press the Silence Alarm button to
2
and pulse rate values have flashing yellow highlights behind them.
2
and pulse rate values are highlighted by flashing yellow boxes.
2
and PR alarm limits as described in
Alarm
Silenced, page 4-12.
Low.
2
Service Manual