Figure 4-11.  Sensor Port - Covidien Nellcor PM10N Service Manual

Portable spo2 patient monitoring system
Hide thumbs Also See for Nellcor PM10N:
Table of Contents

Advertisement

Pinch the ends of the finger sensor to open the sensor to its widest point.
5.
After the monitoring system completes POST, verify the finger sensor LED is brightly lit.
6.
Allow the sensor clip to close slowly.
7.
Verify the LED intensity decreases as the LED approaches the optical sensor.
8.
Open the sensor again and verify that the LED intensity increases.
9.
Repeat step
10.
microprocessor is in proper control of LED intensity.
Leave the monitoring system on for the next test.
11.
Operation with a Live Subject
Use the same finger sensor used in the
with a live subject.
To test using a live subject
Attach the finger sensor to a live subject as recommended in the OxiMax™ pulse oximetry sensor's
1.
instructions for use.
If not already done, press and hold the Power On/Off button to turn the monitoring system on and
2.
verify the monitoring system is operating.
The monitoring system should stabilize on the subject's physiological signal in about 15 to 30 seconds.
3.
Verify the oxygen saturation and pulse rate values are reasonable for the subject.
Alarms and Alarm Silenced
Access these settings from the Service Menu by using the pass code provided by Technical
Services. Reference
To adjust the audio (alarms and alarm silenced) options
Connect the DEC-4 pulse oximetry cable to the monitoring system's sensor port.
1.
Connect the OxiMax™ DS-100A sensor to the DEC-4 cable, then to a finger.
2.
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
3.
Service Manual
7
and verify the intensity continues to decrease. This variation is an indication the
Technical
Services, page 1-8.
LED Excitation
Test, page 4-10 to perform the following test
Figure 4-11.  Sensor Port
Operational and Functional Tests
4-11