Change the brightness to 0. The factory default is 20%.
7.
After 1 minute, verify the following:
8.
The screen of the monitoring system is darkened.
The LED is flashing green.
After pressing any button or when an alarm occurs, verify the following:
9.
The screen of the monitoring system is brightened.
The LED lights green.
Power Saving
Power Saving turns the screen off after 10 minutes of inactivity.
Power Saving functions when no alarm condition exists and the buttons are not pressed.
To test the Power Saving
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system off.
1.
Connect the DEC-4 pulse oximetry cable to the monitoring system's sensor port.
2.
Connect the OxiMax™ DS-100A sensor to the DEC-4 cable, then to a finger.
3.
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
4.
After 10 minutes, verify the following:
5.
The screen of the monitoring system is turned off.
The LED is flashing at 2.5-second intervals.
After pressing any button except the Power On/Off or Silence Alarm button, verify the following:
6.
The brightness of the screen is set to the desired brightness.
The LED lights green.
Auto Power Off
Auto Power Off turns the monitoring system off after the specified period of inactivity.
Auto Power Off functions when no alarm condition exists, the buttons are not pressed for the
specified period, the sensor is disconnected, and the sensor signal is lost.
To test Auto Power Off
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
1.
Access the Service Menu.
2.
Service Manual
Operational and Functional Tests
4-25