Figure 4-24.  Sensor Off Alarm - Covidien Nellcor PM10N Service Manual

Portable spo2 patient monitoring system
Hide thumbs Also See for Nellcor PM10N:
Table of Contents

Advertisement

Observe the monitoring screen.
10.
Verify the following:
11.
The SpO
The High priority alarm sounds.
The message "Sensor Disconnect" appears.
To change the Alarm Priority Setting for Sensor Off
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system off.
1.
Connect the DEC-4 pulse oximetry cable to the monitoring system's sensor port.
2.
Connect the OxiMax™ DS-100A sensor to the DEC-4 cable, then to a finger.
3.
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
4.
When the monitoring system is powered on, disconnect the sensor from the finger.
5.
Verify that the Sensor Off alarm occurs.
6.
Access the Service Menu.
7.
In the Service Menu, select the Sensor Alarm Priority Settings menu.
8.
Set the Sensor Off priority to High.
9.
Go to the monitoring screen.
10.
Service Manual
and pulse rate values are highlighted by flashing red boxes.
2
Figure 4-24.  Sensor Off Alarm
Operational and Functional Tests
4-23