Verify the following:
10.
Active audio alarm.
a.
Flashing %SpO
b.
BPM indication in the range of 57 to 63.
c.
Pulse waveform of approximately 1/2-inch P-T-P amplitude.
d.
Press the SRC-MAX MODULATION selection button. The SRC-MAX MODULATION LED lights. The
11.
monitoring system's pulse waveform increases in amplitude and then stabilizes at a P-T-P amplitude
of approximately 1 inch.
Service Manual
Figure 4-43. %SpO
indication in the range of 73 to 77.
2
of 75 with High Light Condition
2
Operational and Functional Tests
4-43