Maintenance, Overhaul, and Software Upgrade
To replace the emergency relief diaphragm, refer to Figure 2-5.  and Figure 2-6.  , and perform the
following steps:
Access the inhalation outlet assembly by removing the lower right front panel (see Figure 2-5.  ).
1.
Remove the inhalation outlet assembly by removing the two retaining screws (see Figure 2-6.  ).
2.
Remove the four screws from the emergency valve cap to expose the emergency relief valve.
3.
With your finger, pull the emergency relief diaphragm (DIA1800M) out of the inhalation outlet block.
4.
Replace the emergency relief diaphragm.
5.
Reinstall the components and assembly by reversing steps 1-4.
6.

In-Line Disk Filters and Restrictor with Filter

2.7.5
To replace the in-line disk filters and restrictor with filter:
Remove the six screws securing the ventilator's cover.
1.
Remove the cover.
2.
Using the pneumatic diagram in Appendix C, locate and replace the four disk filters (FLT2102P).
3.
Note:
The FLT2102P filters have luer lock connections and can be replaced by twisting the adapters while holding
the filter in your hand.
Using the pneumatic diagram in Appendix C, locate and replace the restrictor with filter (RES2102P).
4.
When replacing it, the arrow on the restrictor should face the direction of flow.
2-10
Figure 2-6.  Emergency Relief Diaphragm Replacement
1
Outlet retaining screws
2
Inhalation adapter
3
Inhalation outlet block
4
Emergency relief diaphragm (DIA1800M)
5
Emergency valve cap
Service Manual