Modification and Testing

Required Equipment

4.2
SpO
SpO
SpO
Stop watch
Note:
Contact Covidien Technical Services for pricing and ordering information of the required equipment.
Reference
Technical

System Performance Tests

4.3
Caution:
If using the PCSync™* utility to change settings for the monitoring system, make sure all the
settings on the PCSync™* screen are correct before clicking Send. Otherwise, unexpected settings
will result.

Power-On Self-Test (POST)

4.3.1
To check the power-on self-test (POST)
With the monitoring system off, press and hold the Power On/Off button for approximately 1 second.
1.
While the monitoring system performs power-on self-test (POST), a progress bar appears at the
bottom of the display.Verify that the following steps occur:
The product name and software version number are displayed on the POST screen.
a.
When POST completes successfully, one pass tone sounds, and the monitoring system displays
b.
the main monitoring screen.
4-2
Table 4-1.  Required Test Equipment
Equipment
sensor (durable)
2
extension cable
2
simulator
2
Services, page 1-8.
Description/use
DS-100A Durasensor™ adult finger clip sensor
DEC-4
Nellcor™ model SRC-MAX
Manual or electronic, for alarm silenced and alarm reminder
intervals
Service Manual