Modification and Testing
Sensor Alarm Priority Settings
Access the Sensor Alarm Priority Settings menu to change the alarm priority to Low, Medium, or
High for:
Sensor Disconnect alarm
Sensor Off alarm
Sensor Failure alarm
To change the Alarm Priority Setting for Sensor Disconnect
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system off.
1.
Connect the DEC-4 pulse oximetry cable to the monitoring system's sensor port.
2.
Connect the OxiMax™ DS-100A sensor to the DEC-4 cable, then to a finger.
3.
Press and hold the Power On/Off button for approximately 1 second to turn the monitoring system on.
4.
When the SpO
5.
system's sensor port.
Verify that the Sensor Disconnect alarm occurs.
6.
Access the Service Menu.
7.
In the Service Menu, select the Sensor Alarm Priority Settings menu.
8.
Set the Sensor Disconnect priority to High.
9.
4-22
saturation and pulse rate are displayed, disconnect the sensor from the monitoring
2
Figure 4-23.  Sensor Disconnect Alarm Priority Setting
Service Manual