Figure 3-13.  Firmware Upgrade Process - Covidien Nellcor PM10N Service Manual

Portable spo2 patient monitoring system
Hide thumbs Also See for Nellcor PM10N:
Table of Contents

Advertisement

If the Resource file needs to be updated, click Resource Update and select the appropriate .res
8.
file.
When the file is selected, the resource upgrade procedure begins automatically.
Upon completion of the resource update, the Complete message is displayed in the PCSync
utility.
If the Resource file does not need to be updated, skip this step.
Click the Firmware Update button and select the appropriate .bin file.
9.
When the file is selected, the firmware update begins automatically.
Upon completion of the firmware update, the Complete message is displayed in the PCSync
utility.
The screen shown in
monitoring system is automatically reset.
Verify the upgrade is complete by checking the POST screen and firmware version.
10.
Service Manual
Figure 3-13
is displayed as the monitoring system is updated. Then the
Figure 3-13.  Firmware Upgrade Process
Firmware Upgrade
3-13