Remove The Batteries; Replace The Batteries - Covidien Nellcor PM10N Service Manual

Portable spo2 patient monitoring system
Hide thumbs Also See for Nellcor PM10N:
Table of Contents

Advertisement

The monitoring system comes with four AA lithium batteries. You can replace depleted batteries
with lithium or alkaline batteries. By default, the monitoring system uses lithium batteries. If
replacing with alkaline batteries, change the Battery Type setting through the Service Menu to
optimize power usage. Reference

Remove the Batteries

6.4.1
Turn the monitoring system off.
1.
Remove the battery cover (3) on the rear housing.
2.
Remove the four AA batteries (2).
3.

Replace the Batteries

6.4.2
Caution:
Use only AA size batteries.
Reference
Figure
Insert four AA batteries. Ensure the correct orientation of each battery (+/-).
1.
Close the battery cover.
2.
Service Manual
Service Menu
Figure 6-3.  Battery Replacement
6-3:
Overview, page 2-1.
Battery Replacement
6-5