Output Off Register

The Output Off Enable Register allows you to determine which conditions and events in the LDX-3200 Series
Precision Current Source can cause the current source output to be turned off. This register is configured in a manner
which is similar to the status reporting registers. However, the register output is not reported in the Status Byte
Register. Rather, it sets the hardware which controls the output switching. The events and conditions which may be
set to cause the current source output to be turned off are shown in Figures 3.3.
The default (factory) settings for this register are shown in Table 3.3
(Power-On Status Clear) command.
Table 3.3 LDX-3200 Series Default Settings for Output Off Registers
OPERATING IN REMOTE CONTROL
CHPT 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
LDX-3200 Series OUTPUT OFF REGISTER
0- disabled
8- N/A
1- disabled
9- disabled
2- N/A
10- disabled
3- enabled
11- enabled
4- N/A
12- N/A
5- N/A
13- N/A
6- N/A
14- N/A
7- N/A
15- N/A
PAGE 9
.
These settings are not effected by the *PSC