ILX Lightwave LDX-3200 Series User Manual page 92

Precision current source
Table of Contents

Advertisement

n Front Panel
n Remote
The LASer:SET:MDP? query returns the laser monitor PD power set point value (when CALMD [CAL PD] is
not zero), in mW.
SYNTAX DIAGRAM
-where the response is the constant P
POINTS OF
INTEREST
In local operation, the MDP set point is read by selecting P mode, then pressing and holding
down the SELECT switch in the DISPLAY area of the front panel. When the mW indicator is
blinking, the display will show the MDP set point.
This set point is used in constant MDP mode only.
EXAMPLES
"LAS:Set:MDP?" -response: 10.0, means the laser monitor PD feedback set point is 10.0 mW
(CALPD > 0).
"Laser:Set:MDP?" -response: 25.0, means the laser monitor PD feedback set point is 25.0 mW
(CALPD > 0).
¨ Front Panel
n Remote
The LASer:STEP command is used to increment or decrement the selected laser control mode set point by the
given amount, when used with the LASer:INC or LASer:DEC command.
SYNTAX DIAGRAM
:
PARAMETERS
An integer value of the step amount, in the range 1 to 9999.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
MDP
set point, in mW.
PD
:
LAS
STEP
PAGE 47
LASer:SET:MDP?
:
SET
P
?
LASer:STEP
<white
<nrf
space>
value>
CHPT 4