ILX Lightwave LDX-3200 Series User Manual page 96

Precision current source
Table of Contents

Advertisement

POINTS OF
INTEREST
The response data will be a 16-byte long string. If there is no previously stored message, the
response will be "
EXAMPLES
"MES?" -response: "Test 3
"Message?" -response: "This is a test. ", means the previously stored message was "This is a
test."
¨ Front Panel
n Remote
The RADix command allows the programmer to select the radix type for status, condition, and event query
response data. Decimal, binary, hexadecimal, and octal are allowed.
SYNTAX DIAGRAM
PARAMETERS
Character program data is expected, as shown above.
POINTS OF
INTEREST
DECimal is the default type. Only the first three letters of the words decimal, hexadecimal,
binary, or octal are required.
When the RADIX is selected, all status, condition, and event queries will return values in the
new radix.
In the cases where the radix is not DECimal, the flexible numeric type <nrf value> (as shown in
the Command Reference diagrams) will be replaced by HEX, BIN, or OCT representation.
All of the above radixes may be used to enter program data at any time, without the need for
issuing the RADix command. The proper prefix must also be used with Hex (#H), binary (#B), or
octal (#Q).
This command may be useful for setting up status reporting blocks. The bit-wise status
representation may be more easily read in BIN, HEX, or OCT.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
", all spaces.
", means the previously stored message was "Test 3".
<white
:
RAD
space>
DEC
BIN
HEX
PAGE 51
RADix
OCT
CHPT 4