ILX Lightwave LDX-3200 Series User Manual page 70

Precision current source
Table of Contents

Advertisement

¨ Front Panel
n Remote
The LASer:ENABle:EVEnt? query returns the value of the status event enable register of the LASER
operations.
SYNTAX DIAGRAM
:
-where the response is the sum of the following:
1 - LASER Current Limit
2 - LASER Voltage Limit
4 - N/A
8 - Power Limit
16 - Interlock State Changed
32 - N/A
64 - Compliance Voltage Limit
128 - Open Circuit
POINTS OF
INTEREST
The enabled LASER events can be set by using the LASer:ENABle:EVEnt command. The
LASER event status can be monitored by the LASer:EVEnt? query.
EXAMPLES
"LAS:ENAB:EVE?" -response: 1040, means that the Output on/off state change and Interlock
changed LASER events will be reported (in summarized form) to the status byte register (bit 2).
"Radix Hex; Las:Enab:Eve?" -response: #HFDB, means that all of the above events will be
reported (in summarized form) to the status byte register (bit 2).
¨ Front Panel
n Remote
The LASer:ENABle:OUTOFF command sets the status outoff enable register of the LASER operations (things
which will turn the LASER output off).
SYNTAX DIAGRAM
:
LAS
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
LAS
ENAB
256 - Output is Shorted
512 - Output Changed to be In/Out of Tolerance
1024 - Output On/Off State
2048 - New Measurements Taken
4096 - N/A
8192 - N/A
16384 - N/A
32768 - N/A
:
ENAB
PAGE 25
LASer:ENABle:EVEnt?
:
EVE
LASer:ENABle:OUTOFF