ILX Lightwave LDX-3200 Series User Manual page 76

Precision current source
Table of Contents

Advertisement

¨ Front Panel
n Remote
The LASer:LDV? query returns the value of the measured laser voltage.
SYNTAX DIAGRAM
-where the response is the measured laser output voltage, for either low or high bandwidth
modes.
POINTS OF
INTEREST
This measurement is updated approximately once every 600 msec.
EXAMPLES
"LAS:ldv?" -response: 3.03, means the measured laser output voltage is 3.03 volts.
"Laser:LDV?" -response: 1.0, means the measured laser output voltage is 1.000 volt.
n Front Panel
n Remote
The LASer:LIMit:I1 command sets the LASER current limit value for the 100 mA range on the LDX-3210. (Not
applicable for LDX-3220).
SYNTAX DIAGRAM
:
LAS
PARAMETERS
An <nrf value> which represents the LASER limit current, in mA.
POINTS OF
INTEREST
The current limit is in effect in all modes of operation (ILBW, IHBW, or P). The current limit in
use depends on the LASER output range selection.
EXAMPLES
"LAS:LIM:I1 80" -action: the LASER current limit is set to 80 mA.
":Laser:Limit:I1 60" -action: the LASER current limit is set to 60 mA.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
:
:
LIM
PAGE 31
LASer:LDV?
?
LDV
LASer:LIMit:I1
<white
<nrf
I1
value>
space>
CHPT 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldx-3220Ldx-3210

Table of Contents