ILX Lightwave LDX-3200 Series User Manual page 75

Precision current source
Table of Contents

Advertisement

n Front Panel
n Remote
The LASer:LDI? query returns the value of the measured laser current.
SYNTAX DIAGRAM
-where the response is the most recent laser current measurement.
POINTS OF
INTEREST
Response is the measured laser output current, for either low or high bandwidth modes.
This measurement is updated approximately once every 600 msec.
EXAMPLES
"LAS:ldi?" -response: 30.0, means the measured laser output current is 30.0 mA.
"Laser:LDI?" -response: 100.0, means the measured laser output current is 100.0 mA.
¨ Front Panel
n Remote
The LASer:LDV command sets the laser voltage for calibration of the laser voltage measurement.
SYNTAX DIAGRAM
:
PARAMETERS
An <nrf value> which represents the (laser) output voltage, in volts.
POINTS OF
INTEREST
Set point is the same for both low and high bandwidth output modes and is entered only for
LDV calibration.
EXAMPLES
"Las:LDV 4" -action: enters a value of 4.000 volts.
"LAS:ldv 1.025" -action: enters a value of 1.025 volts.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
:
LAS
PAGE 30
LASer:LDI?
I
?
LDI
LASer:LDV
<white
<nrf
LDV
space>
value>
CHPT 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldx-3220Ldx-3210

Table of Contents