ILX Lightwave LDX-3200 Series User Manual page 58

Precision current source
Table of Contents

Advertisement

¨ Front Panel
n Remote
The LASer:COND? query returns the value of the status condition register of the LASER operations.
SYNTAX DIAGRAM
-where the response is the sum of the following:
1 - LASER Current Limit
2 - LASER Voltage Limit
4 - N/A
8 - Power limit
16 - Interlock disabled
32 - N/A
64 – Compliance Voltage Limit
128 - Open circuit
POINTS OF
INTEREST
The LASER conditions which are reported to the status byte are set via the
LASer:ENABle:COND command.
The Open circuit condition is only present while the LASER output is on, and when the
hardware detects this condition, it will turn the LASER output off. Therefore, the Open Circuit
condition is fleeting and may be missed via the LAS:COND? query. Therefore, the user should
test for the Open Circuit Event via the LAS:EVEnt? query.
The LASER condition status is constantly changing, while the event status is only cleared
when the event status is read or the *CLS command is issued.
EXAMPLES
"LAS:COND?" -response: 513, means that the LASER limit current and out of tolerance LASER
conditions currently exist.
"Radix Hex; Laser:Cond?" -response: #H108, means that the LASER Output shorted (output is
off) and Power limit conditions currently exist.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
256 - Output is shorted
512 - Output is outside tolerance limit
1024 - Output on/off state
2048 - Ready for calibration data state
4096 - N/A
8192 - N/A
16384 - N/A
32768 - N/A
PAGE 13
LASer:COND?
?
COND
CHPT 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldx-3220Ldx-3210

Table of Contents