POINTS OF
INTEREST
After this query is issued and a response of 1 is received, the LDX-3200 Series Precision
Current Source will be ready for the user to enter a current value via the LASer:LDI command
(see Chapter 6).
EXAMPLES
"LASer:CAL:LDI?" -response: 1, means the LDX-3200 Series Precision Current Source is ready
for the user to enter a current value via the LASer:LDI command.
"LASer:Cal:LdI?" -response: 0, means the LDX-3200 Series Precision Current Source is not yet
ready for the user to enter a LASER current value.
n Front Panel
n Remote
The LASer:CAL:LDV command is used to enter the LASER voltage measurement calibration mode.
SYNTAX DIAGRAM
PARAMETERS
None.
POINTS OF
INTEREST
After this command is issued, the LDX-3200 Series Precision Current Source will allow
calibration of the laser voltage measurement. This procedure is outlined in Chapter 6.
The LASer:CAL:LDV? query may be used to determine if the LDX-3200 Series Precision Current
Source is ready for the user to enter a value.
EXAMPLES
"Las:CAL:Ldv" -action: the LDX-3200 Series Precision Current Source enters calibration mode
for LASER voltage.
"Las:Cal:LDV" -action: the LDX-3200 Series Precision Current Source enters calibration mode
for LASER voltage.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
PAGE 10
LASer:CAL:LDV
:
CAL
LDV
CHPT 4