ILX Lightwave LDX-3200 Series User Manual page 88

Precision current source
Table of Contents

Advertisement

POINTS OF
INTEREST
There is a 2-second delay after the output is turned on before the output shorting relay is
opened. Following this, the output current ramps up to the current set point slowly to protect
the laser. Therefore, the time to turn the output on and reach the set point current varies from 2
to about 3 seconds.
After the output is turned on, it may be useful to wait until the output is stable (within
tolerance) before performing further operations, but it is not necessary. When the LASER
output is off, it is safe to connect or disconnect devices to the LASER output terminals.
When the LASER output is off, an internal short is placed across the output terminals.
EXAMPLES
"LAS:I 20; LAS:OUT ON" -action: sets the laser output current to 20 mA and then turns the
output on.
n Front Panel
n Remote
The LASer:OUTput? query returns the status of the laser output switch.
SYNTAX DIAGRAM
-where the response is 0 or 1.
POINTS OF
INTEREST
Although the status of the switch is on, the output may not have reached the set point value.
EXAMPLES
"Las:OUT?" -response: 0, means that the output switch is disabled, devices may be safely
disconnected or connected at the LASER output terminals.
"LAS:OUT?" -response: 1, means that the LASER output switch is enabled, LASER output is
present.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
PAGE 43
LASer:OUTput?
?
OUT
CHPT 4