Output Off Register; Figure 4.7 Laser Output Off Register; Table 4.9 Default Output Off Register Settings - ILX Lightwave LDX-36000 Series User Manual

High power laser diode current source
Table of Contents

Advertisement

4
C H A P T E R

Output Off Register

The Output Off Enable Register determines which conditions and events can
cause the laser current output to be turned off. This register is configured in a
manner which is similar to the status reporting registers. However, its output is not
reported in the Status Byte Register. Rather, it is directly tied to hardware which
controls the output switching. The events and conditions which may be set to
cause the laser output to be turned off are shown in Figure 4.7. The default
(factory) settings are shown in Table 4.9. The settings preceeded by an asterisk
(*) are hard-coded and cannot be changed. These settings are not affected by the
*PSC (Power-On Status Clear) command.

Table 4.9 Default Output Off Register Settings

Bit Reference
00 - Disabled
01 - *Enabled
02 - Enabled
03 - *Enabled
04 - *Enabled
05 - *Enabled
06 - Disabled
07 - Disabled
62
LDX-36000 Series
R E M O T E O P E R A T I O N
Status Reporting

Figure 4.7 Laser Output Off Register

08 - *Disabled
09 - Disabled
10 - *Enabled
11 - *Enabled
12 - *Enabled
13 - *Enabled
14 - *Enabled
15 - *Enabled

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents