ILX Lightwave LDX-3200 Series User Manual page 69

Precision current source
Table of Contents

Advertisement

¨ Front Panel
n Remote
The LASer:ENABle:EVEnt command sets the status event enable register of the LASER operations. These
events are summarized in bit 2 of the status byte register.
SYNTAX DIAGRAM
:
LAS
PARAMETERS
An <nrf value> whose sum represents the bits which are enabled:
1 - LASER Current Limit
2 - LASER Voltage Limit
4 - N/A
8 - Power Limit
16 - Interlock Changed State
32 - N/A
64 - Compliance Voltage Limit
128 - Open Circuit
POINTS OF
INTEREST
The enabled LASER events can be read by using the LASer:ENABle:EVEnt? query. The LASER
event status can be monitored by the LASer:EVEnt? query.
The enable registers normally retain their values at power-up (as they were at power-down)
unless the power-on status clear flag is set true (for more information see the *PSC definition, in
the IEEE488.2 specification).
EXAMPLES
"LAS:ENAB:EVENT 136" -action: enables the LASER status event register so that the Open
circuit and Power limit events will be reported (in summarized form) to the status byte (bit 2).
"Laser:Enable:Event #H0FDB" -action: enables the LASER status event register so all of the
above events will be reported (in summarized form) to the status byte (bit 2).
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
ENAB
256 - Output is Shorted
512 - Output Changed to be In/Out of Tolerance
1024 - Output On/Off State Changed
2048 - New Measurements Taken
4096 - N/A
8192 - N/A
16384 - N/A
32768 - N/A
PAGE 24
LASer:ENABle:EVEnt
<white
<nrf
EVE
value>
space>