ILX Lightwave LDX-3200 Series User Manual page 98

Precision current source
Table of Contents

Advertisement

¨ Front Panel
n Remote
The TERM command allows the programmer to change the default (IEEE488.2 standard) terminator to include
the carriage return.
SYNTAX DIAGRAM
PARAMETERS
An <nrf value>, 0 = FALSE, non-zero = TRUE
POINTS OF
INTEREST
An altered terminator will be in the form <CR><NL><^END>. This technically takes the LDX-
3200 Series Precision Current Source out of IEEE488.2 specification, but may be done for
convenience when using non-standard GPIB interfaces. This termination will be sent with all
output until the "TERM 0" command is sent or the LDX-3200 Precision Current Source is
powered off.
EXAMPLES
"Term 1" -temporarily sets <CR><NL><^END> as the output terminator.
"Term 0" -sets the IEEE488.2 standard <NL><^END> terminator. This is the LDX-3200 Series'
default value.
¨ Front Panel
n Remote
The TERM? query allows the programmer to determine whether the default (IEEE488.2 standard) terminator
has been altered to include a carriage return.
SYNTAX DIAGRAM
-where the response is 0 or 1.
POINTS OF
INTEREST
An altered terminator will be in the form <CR><NL><^END>. This termination will be sent with
all output until the "TERM 0" command is sent or the LDX-3200 Series Precision Current Source
is powered off.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
<white
:
TERM
space>
:
TERM
PAGE 53
TERM
<nrf
value>
TERM?
?
CHPT 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldx-3220Ldx-3210

Table of Contents