ILX Lightwave LDX-3200 Series User Manual page 72

Precision current source
Table of Contents

Advertisement

-where the response is the sum of the following:
1 - LASER Current Limit
2 - LASER Voltage Limit
4 - N/A
8 - Power Limit (With Output On)
16 - N/A
32 - N/A
64 - N/A
128 - N/A
POINTS OF
INTEREST
The enabled LASER events can be set by using the LASer:ENABle:OUTOFF command. The
LASER output status can be monitored by the LASer:EVEnt? query.
EXAMPLES
"LAS:ENAB:OUTOFF?" -response: 1, means that the Current Limit Condition will cause the
LASER output to be turned off.
"Radix Hex; Las:Enab:Eve?" -response: #H20B, means that all of the above conditions will
cause the LASER output to be turned off.
¨ Front Panel
n Remote
The LASer:EVEnt? query returns the value of the status event register of the LASER operations.
SYNTAX DIAGRAM
-where the response is the sum of the following:
1 - LASER Current Limit
2 - LASER Voltage Limit
4 - N/A
8 - Power Limit
16 - Interlock Disabled
32 - N/A
64 - Compliance Voltage Limit
128 - Open Circuit
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
256 - N/A
512 - Output is Out of Tolerance
1024 - N/A
2048 - N/A
4096 - N/A
8192 - N/A
16384 - N/A
32768 - N/A
:
:
LAS
256 - Output is Shorted
512 - Output Changed to be In/Out of Tolerance
1024 - Output On/Off State Changed
2048 - Measurement Ready
4096 - N/A
8192 - N/A
16384 - N/A
32768 - N/A
PAGE 27
LASer:EVEnt?
?
EVE
CHPT 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldx-3220Ldx-3210

Table of Contents