ILX Lightwave LDX-3200 Series User Manual page 84

Precision current source
Table of Contents

Advertisement

POINTS OF
INTEREST
The LASER modulation control switches the front panel BNC in and out of the LAS control
circuit. At power-up or after reset, modulation is enabled by default.
EXAMPLES
"LAS:MDLN?" -response: 1, means that the LAS modulation input is enabled.
"las:mdln?" -response: 0, means that the LAS modulation input is disabled.
n Front Panel
n Remote
The LASer:MDP command sets the value of the optical power set point, in mW, if the CALMD (CAL PD)
responsivity is greater than 0.
SYNTAX DIAGRAM
:
PARAMETERS
An <nrf value> which represents the photodiode feedback power, in mW.
POINTS OF
INTEREST
If the CALMD (CAL PD) parameter is set to 0, the LAS:MDP value will not be used. In this
case, the measured MDP would be invalid, and the MDI (I
EXAMPLES
"Las:Mdp 40" -action: The LASER output is controlled so that the photodiode feedback
power remains constant at 40 mW.
"Laser:MDP 200" -action: The LASER output is controlled so that the photodiode feedback
power remains constant at 200 mW.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
MDP
:
LAS
PAGE 39
LASer:MDP
<white
space>
P
) set point would be used.
PD
<nrf
value>
CHPT 4