ILX Lightwave LDX-3200 Series User Manual page 85

Precision current source
Table of Contents

Advertisement

n Front Panel
n Remote
The LASer:MDP? query returns the value of the laser photodetector power measurement, in mW.
SYNTAX DIAGRAM
-where the response is the most recent LASER power measurement.
POINTS OF
INTEREST
The response is in mW. The response is valid, even when the unit is not in constant P mode.
This measurement is updated approximately once every 600 ms ec.
EXAMPLES
"Las:Mode:MDP; Las:Calmd 460; Las:MDp?"
photodetector power. This feedback is controlling the laser current output.
"LAS:MODE:IHBW; LAS:MDP?" -response: 20.0, means 20 mW of photodetector power, but
photodiode monitor current is not controlling the laser output current.
n Front Panel
n Remote
The LASer:MODE? query returns the selected laser control mode.
SYNTAX DIAGRAM
-where the response is the LASER control mode.
POINTS OF
INTEREST
I
mode is the same as I mode (low bandwidth), except that the output low bandpass filter is
HBW
disabled in I
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
:
:
LAS
mode.
HBW
PAGE 40
LASer:MDP?
MDP
?
P
-response:
100.0, means 100 mW of
LASer:MODE?
?
MODE
CHPT 4