Key Operator Instructions

Verifying the Replenishment Rates

5–50
8. Press DEV REP VOLUME to change the Developer Replenishment volume, or
FIX REP VOLUME to change the Fixer Replenishment volume.
DEV REP
FIX REP
VOLUME
VOLUME
H186_8550AC
9. To change the replenishment volume press:
a. ∆ to increase
b. ∇ to decrease
c. DEFAULT SETTING to return to the factory setpoint
d. CANCEL REQUEST to return to the last value set
H186_8551AC
10. Press DONE/RETURN several times to store the entry, and to return to the
Processor Main Menu.
11. Do Steps 8 - 10 for the other solution.
Warning
.
• Gloves and protective eye wear must be worn when working with the processor
chemistry, especially when mixing the chemistry.
• See the "Chemical Hazards" warning on Page 1–9.
1. Gently pull out the Processor from the Multiloader as described on Page 4–9, and
remove the Wet Section Cover.
DONE/
RETURN
DEFAULT
CANCEL
DONE/
SETTING
REQUEST
RETURN
16FEB01 – 7E6174