Replenishment Pump - Kodak X-Omat 2000 Service Manual

Hide thumbs Also See for X-Omat 2000:
Table of Contents

Advertisement

Warning
Dangerous Voltage
[8] Connect the main power.
Caution
Do not operate the RECIRCULATION PUMP if the TANKS are empty.
[9] Actuate the RECIRCULATION PUMP.
[10] Check for:
• agitation of the solution in the TANK
• leakage

REPLENISHMENT PUMP

EXIT
PANEL
PUMP
COVER
DRYER TEMPERATURE CONTROL
KNOB
2 REPLENISHMENT PUMPS
2 WASHERS
2 LOCK WASHERS
2 SCREWS
[4] Disconnect:
• MOTOR LEAD TB3-6
• MOTOR LEAD TB3-2
• GROUND WIRE
[5] Install PINCH CLAMPS on the TUBING to the REPLENISHMENT PUMP.
7C8772 – 12JAN99
TOP COVER
2 SIDE PANELS
H172_1028ACB
H172_1028AC
Adjustments and Replacements
Warning
Dangerous Voltage
[1] Disconnect the main power.
[2] Disconnect the water supply to the PROCESSOR.
[3] Remove:
• TOP COVER
• 2 SIDE PANELS
• RECEIVING BIN
• DRYER TEMPERATURE CONTROL
KNOB
• PUMP COVER
• EXIT PANEL
REPLENISHER TUBE LINES
H172_1057BCA
H172_1057BC
43