Section 3: Periodic Maintenance
Procedure and Schedule
Important
The Periodic Maintenance procedures include additional procedures which are not in the OPERATOR MANUAL,
Publication No. 7C8770. For optimum operation of the PROCESSOR, it is important to do these procedures.
[1] Process 5 sheets of unprocessed film.
[2] Check the film for artifacts.
[3] Do all of the Monthly Maintenance Procedures first.
[4] Do the procedure for lubrication on Page 53.
[5] Check the specifications of the site.
[6] Process 5 sheets of film.
[7] Check the film for artifacts.
Monthly Maintenance Procedures
DETECTOR SWITCHES
[1] Check that the DETECTOR SWITCHES operate the REPLENISHMENT PUMP.
[2] Adjust the DETECTOR SWITCHES if necessary. Use the procedure on Page 5.
DETECTOR CROSSOVER AY
[1] Use a damp, lint-free CLOTH or a SPONGE to clean the DETECTOR ROLLERS.
[2] Check:
• ROCKER ARM moves freely
• ROCKER ARM does not have corrosion
• DETECTOR CROSSOVER AY for squareness
[3] Clean the ROCKER ARM if it does not move freely.
[4] Install a new ROCKER ARM if it has corrosion.
[5] Check the DETECTOR CROSSOVER AY for squareness. Use the procedure on Page 7.
[6] Check that the DETECTOR CROSSOVER AY is correctly seated.
CROSSOVER AYs
[1] Remove the CROSSOVER AYs from the PROCESSOR.
[2] Rinse the CROSSOVER AYs with water.
[3] Use warm water and a soft BRUSH to clean the CROSSOVER AYs, or use:
DEVELOPER SYSTEM CLEANER and
NEUTRALIZER, Catalog No. 843 4615
FIXER/WASH SYSTEM CLEANER,
Catalog No. 139 5110
7C8772 – 12JAN99
Cleaning Solution
Type of CROSSOVER
DEVELOPER/FIXER CROSSOVER
FIXER/WASH CROSSOVER
Periodic Maintenance
45