PLUMBING System Adjustments

Adjusting the Developer Temperature

[1] Remove the TOP COVER.
[2] Wait until the TEMPERATURE READY LIGHT blinks slowly or does not illuminate.
[3] Insert a THERMOMETER into the DEVELOPER TANK on the non-drive side of the PROCESSOR between
the SIDE PLATE of the DEVELOPER RACK and the RACK SUPPORT.
[4] Check that the temperature of the developer is correct for the operating cycle of the PROCESSOR.
100 CIRCUIT BOARD
R2
[5] If the developer temperature is correct, advance to Step 9.
7C8772 – 12JAN99
Cycle
±
Standard
33.3
0.3
±
Rapid
34.4
0.3
Insert THERMOMETER
here
Temperature
°
±
°
C (92.0
0.5
F)
°
±
°
C (94.0
0.5
F)
ELECTRICAL
BOX
Adjustments and Replacements
H172_1006BCB
H172_1006BC
H172_1031BCB
H172_1031BC
29